Merek | Philips |
---|---|
Model | S5-1 |
Kesalahan | Penuaan elemen array, gelembung lensa |
Larutan | Ganti Kristal |
Kemasan rincian | Paket standar |
Merek | Medtronic |
---|---|
Model | 1898200t |
Kesalahan | 1) bantalan korosi dan segel rusak. 2) Motor terjebak, di dalam penuh residu desinfeksi, korosi moto |
Larutan | Bertukar seluruh handpiece kecuali kabel. |
Kemasan rincian | Paket standar |
Merek | Ge |
---|---|
Model | RIC6-12-D |
Kesalahan | Masalah lensa |
Larutan | Ganti lensa, kabel bersih |
Kemasan rincian | Paket standar |
Brand | Karl Storz |
---|---|
Model | 27005BA |
Fault | 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. |
Solution | 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope a |
Packaging Details | Standard Package |
Brand | OLYMPUS |
---|---|
Model | TJF-Q180V |
Fault | 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. |
Solution | Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and rubber. (Nowarranty for CCD) |
Kemasan rincian | Paket standar |
Merek | penyimpanan |
---|---|
Model | 13820pks |
Kesalahan | 1. Kebocoran Rubber; 2. Abrasi; 3. Tidak ada sudut, kurang dari sudut lain |
Larutan | 1. Ganti kabel, pipa kumparan, karet; 2. Sungguh sudut |
Kemasan rincian | Paket standar |
Merek | Olympus |
---|---|
Model | GIF-Q150 |
Kesalahan | 1. dengan gambar; 2. Satu mata bundel cahaya rusak; 3. Tabung Tabung Sedikit Penuaan; 4.Biopsy Tube |
Larutan | Ganti bundel lampu, tabung insert, tabung biopsi, bagian lentur, kabel, tabung helix, tabung pemandu |
Kemasan rincian | Paket standar |
Merek | Olympus |
---|---|
Model | GIF-H180J |
Kesalahan | 1.Image dengan bayangan; 2. Bundel cahaya rusak; 3. Tabung insert dengan tanda gigitan; 4.Biopsy Tub |
Larutan | Ganti CCD, bundel lampu, tabung insert, tabung biopsi, tabung air & air, bagian lentur, kabel, t |
Kemasan rincian | Paket standar |
Brand | Olympus |
---|---|
Model | CHF-P20Q |
Fault | 1.Insert tube cracked; 2.Rubber leakage; 3.Image with black point and cracked; 4.Angulation tight; 5.Biopsy tube aging |
Solution | Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber |
Packaging Details | Standard Package |
Merek | Olympus |
---|---|
Model | BF-XP160F |
Kesalahan | 1. Gambar degumming, warna serius, penuaan CCD, beberapa bintik hitam dan abu -abu; 2. 20% dari bund |
Larutan | Ganti CCD, balok gambar, balok, tabung insert, tabung tang, tulang ular, kawat baja, tabung spiral, |
Kemasan rincian | Paket standar |